Abstract

Read online

The paper considers the problem of distortion of the ECG curve caused by bioimpedance of a patients skin. The authors compare traditional analytical methods of bioimpedance estimation based on equivalent circuits using both basic RLC elements and advanced elements (traditional Warburg, mass dissipative Warburg, etc.) and black box method. Advantage of black box method is shown in the task of estimation of parameters of parasitic filter of the skin in comparison with traditional methods. The paper describes the theoretical base and the measurement experiment directed to estimation of real transition parameters of the correction filter. Obtained digital correction filter was experimentally investigated. The results of the computational experiment with ECG records from PhysioNET base shows the impact of bioimpedance changes to ECG signal form. The proposed solution is intended to implementation in portable heart monitoring systems.

Keywords