Sensors (May 2016)

Use of a Force-Torque Sensor for Self-Calibration of a 6-DOF Medical Robot

  • Ahmed Joubair,
  • Long Fei Zhao,
  • Pascal Bigras,
  • Ilian A. Bonev

DOI
https://doi.org/10.3390/s16060798
Journal volume & issue
Vol. 16, no. 6
p. 798

Abstract

Read online

The aim of this paper is to improve the position accuracy of a six degree of freedom medical robot. The improvement in accuracy is achieved without the use of any external measurement device. Instead, this work presents a novel calibration approach based on using an embedded force-torque sensor to identify the robot’s kinematic parameters and thereby enhance the positioning accuracy. A simulation study demonstrated that our calibration approach is effective, whether or not any measurement noise is present: the position error is improved, inside the robot target workspace, from 12 mm to 0.320 mm, for the maximum values, and from 9 mm to 0.2771 mm, for the mean errors.