Blood Pressure (Dec 2023)

Performance of the Aktiia optical blood pressure measurement device in the elderly: a comparison with double blinded auscultation in different body positions

  • Kenji Theiler,
  • Josep Sola,
  • Aikaterini Damianaki,
  • Alexandre Pfister,
  • Tiago P Almeida,
  • Jérémy Alexandre,
  • Pascale Vermare,
  • Gregoire Wuerzner

DOI
https://doi.org/10.1080/08037051.2023.2281320
Journal volume & issue
Vol. 32, no. 1

Abstract

Read online

AbstractBackground: Accurate blood pressure (BP) measurement is essential for the correct diagnosis and management of hypertension (HTN) especially in the elderly population. As with of all BP devices, the accuracy of cuffless devices must be verified. This study (NCT04027777) aimed to evaluate the performance of a wrist cuffless optical BP device in an elderly population cohort in different body positions with auscultation as the reference measurement.Design and methods Patients aged 65-85 years with different BP categories but without diabetes were recruited. After an initial calibration based on auscultatory measurements, BP estimation from the Aktiia Bracelet (Aktiia SA, Switzerland) were compared to reference double-blinded auscultatory measurements in sitting, standing and lying positions on four separate visits distributed over one month. In the absence of a universal standard for cuffless BP device at the time of the study, modified ISO81060-2 criteria were used for performance analysis.Results Thirty-five participants were included in the analysis fulfilling the inclusion requirements of ISO 81060-2. A total of 469 paired measurements were obtained with overall 83% acceptance rate. Differences (mean ± SD) [Formula: see text] between Aktiia Bracelet and auscultation for systolic BP were -0.26 ± 9.96 mmHg for all body positions aggregated (sitting 1.23 ± 7.88 mmHg, standing -1.81 ± 11.11 mmHg, lying -1.8 ± 9.96 mmHg). Similarly, differences for diastolic BP were -0.75 ± 7.0 mmHg (0.2 ± 5.55 mmHg, -5.35 ± 7.75 mmHg and -0.94 ± 7.47 mmHg, respectively). Standard deviation of the averaged differences per subject for systolic/diastolic BP was 3.8/2.5 mmHg in sitting and 4.4/3.7 mmHg for all body positions aggregated.Conclusions Overall, this study demonstrates a similar performance of the Aktiia Bracelet compared to auscultation in an elderly population in body positions representative of daily activities. The use of more comfortable, non-invasive, and non-occlusive BP monitors during long periods may facilitate e-health and may contribute to better management of HTN, including diagnosis and treatment of HTN, in the elderly.

Keywords