BMC Infectious Diseases (Feb 2020)

Implementation of the FilmArray ME panel in laboratory routine using a simple sample selection strategy for diagnosis of meningitis and encephalitis

  • Susanne Pfefferle,
  • Martin Christner,
  • Martin Aepfelbacher,
  • Marc Lütgehetmann,
  • Holger Rohde

DOI
https://doi.org/10.1186/s12879-020-4904-4
Journal volume & issue
Vol. 20, no. 1
pp. 1 – 9

Abstract

Read online

Abstract Background Infectious meningitis is a serious disease and patient outcome relies on fast and reliable diagnostics. A syndromic panel testing approach like the FilmArray ME can accelerate diagnosis and therefore decrease the time to pathogen specific therapy. Yet, its clinical utility is controversial, mainly because of a remaining uncertainty in correct interpretation of results, limited data on its performance on clinical specimens and its relatively high costs. The aim of this study was to analyze clinical performance of the assay in a real life setting at a tertiary university hospital using a pragmatic and simple sample selection strategy to reduce the overall cost burden. Methods Over a period of 18 months we received 4623 CSF samples (2338 hospitalizations, 1601 individuals). FilmArray ME analysis was restricted to CSF-samples with a high pretest probability of infectious meningitis, e.g. positive Gram-stain, samples in which leukocytes and/or bacteria were evident or urgent suspicion of infection was communicated by clinicians. N = 171 samples matched to our risk criteria and were subjected to FilmArray ME analysis. Those samples were also analyzed by reference methods: culture only (n = 45), PCR only (n = 20) or both methods (n = 106). Results 56/171 (32.75%) were FilmArray ME positive. Bacterial pathogens were detected in 30/56 (53.57%), viral pathogens were detected in 27/56 (48.21%) and yeast DNA was detected in 1/56 (1.79%) of positive samples. Double detection occurred in 2/56 samples. In 52/56 (92.86%) FilmArray ME positive samples, results could be confirmed by the reference assays (sensitivity = 96.30%, specificity =96.58%). Conclusion The FilmArray ME assay is a fast and reliable diagnostic tool for the management of infectious meningitis and can easily be implemented in routine diagnostic workflows. However, correlation of test results and underlying clinical symptoms requires experienced users and the awareness of potentially false negative or false positive results. Moreover, considering the need for antimicrobial susceptibility testing, the use of molecular tests as a stand-alone diagnostic cannot be recommended.

Keywords