Scientific Reports (Jan 2023)

Low field slice-selective ZTE imaging of ultra-short $$T_2$$ T 2 tissues based on spin-locking

  • Jose Borreguero,
  • Fernando Galve,
  • José M. Algarín,
  • José M. Benlloch,
  • Joseba Alonso

DOI
https://doi.org/10.1038/s41598-023-28640-x
Journal volume & issue
Vol. 13, no. 1
pp. 1 – 12

Abstract

Read online

Abstract Magnetic Resonance Imaging of hard biological tissues is very challenging due to small proton abundance and ultra-short $$T_2$$ T 2 decay times, especially at low magnetic fields, where sample magnetization is weak. While several pulse sequences, such as Ultra-short Echo Time (UTE), Zero Echo Time (ZTE) and SWeep Imaging with Fourier Transformation (SWIFT), have been developed to cope with ultra-short lived MR signals, only the latter two hold promise of imaging tissues with sub-millisecond $$T_2$$ T 2 times at low fields. All these sequences are intrinsically volumetric, thus 3D, because standard slice selection using a long soft radio-frequency pulse is incompatible with ultra-short lived signals. The exception is UTE, where double half pulses can perform slice selection, although at the cost of doubling the acquisition time. Here we demonstrate that spin-locking is a versatile and robust method for slice selection for ultra-short lived signals, and present three ways of combining this pulse sequence with ZTE imaging of the selected slice. With these tools, we demonstrate slice-selected 2D ex vivo imaging of the hardest tissues in the body at low field (260 mT) within clinically acceptable times.