Scientific Reports (Jul 2021)

A new approach to precise mapping of local temperature fields in submicrometer aqueous volumes

  • Alexey M. Romshin,
  • Vadim Zeeb,
  • Artem K. Martyanov,
  • Oleg S. Kudryavtsev,
  • Dmitrii G. Pasternak,
  • Vadim S. Sedov,
  • Victor G. Ralchenko,
  • Andrey G. Sinogeykin,
  • Igor I. Vlasov

DOI
https://doi.org/10.1038/s41598-021-93374-7
Journal volume & issue
Vol. 11, no. 1
pp. 1 – 7

Abstract

Read online

Abstract Nanodiamonds hosting temperature-sensing centers constitute a closed thermodynamic system. Such a system prevents direct contact of the temperature sensors with the environment making it an ideal environmental insensitive nanosized thermometer. A new design of a nanodiamond thermometer, based on a 500-nm luminescent nanodiamond embedded into the inner channel of a glass submicron pipette is reported. All-optical detection of temperature, based on spectral changes of the emission of “silicon-vacancy” centers with temperature, is used. We demonstrate the applicability of the thermometric tool to the study of temperature distribution near a local heater, placed in an aqueous medium. The calculated and experimental values of temperatures are shown to coincide within measurement error at gradients up to 20 °C/μm. Until now, temperature measurements on the submicron scale at such high gradients have not been performed. The new thermometric tool opens up unique opportunities to answer the urgent paradigm-shifting questions of cell physiology thermodynamics.