BioTechniques (Aug 1997)

Long-Range and Highly Sensitive DNase I Footprinting by an Automated Infrared DNA Sequencer

  • Masayuki Machida,
  • Hiroshi Kamio,
  • David Sorensen

DOI
https://doi.org/10.2144/97232st08
Journal volume & issue
Vol. 23, no. 2
pp. 300 – 303

Abstract

Read online

We have shown that an automated DNA sequencer is applicable to fluorescencebased detection of fragments in DNase I footprinting. We demonstrated the potential of long-range and highly sensitive DNase I footprinting taking advantage of an infrared- fluorescence automated DNA sequencer. Footprints of human transcription factor Sp1 were reproducibly detected ranging approximately between 100 and 750 bp on both strands of an 895-bp DNA fragment in a single electrophoresis run. We developed techniques in data collection and subsequent image processing for highly sensitive detection. Less than 0.1 footprinting unit (fpu; approximately 4.5 ng) of Sp1 was detected using 3.1 fmol of a 512-bp DNA fragment. This is greater than 10-fold increase in sensitivity over what has previously been reported by visible dye fluorescence DNA sequencers. This method will be very important in systematic analysis of transcription regulatory regions and in largescale analysis of the transcription process.