Measurement Science Review (Oct 2014)

Influence of Wilbraham-Gibbs Phenomenon on Digital Stochastic Measurement of EEG Signal Over an Interval

  • Sovilj P.,
  • Milovanović M.,
  • Pejić D.,
  • Urekar M.,
  • Mitrović Z.

DOI
https://doi.org/10.2478/msr-2014-0037
Journal volume & issue
Vol. 14, no. 5
pp. 270 – 278

Abstract

Read online

Measurement methods, based on the approach named Digital Stochastic Measurement, have been introduced, and several prototype and small-series commercial instruments have been developed based on these methods. These methods have been mostly investigated for various types of stationary signals, but also for non-stationary signals. This paper presents, analyzes and discusses digital stochastic measurement of electroencephalography (EEG) signal in the time domain, emphasizing the problem of influence of the Wilbraham-Gibbs phenomenon. The increase of measurement error, related to the Wilbraham-Gibbs phenomenon, is found. If the EEG signal is measured and measurement interval is 20 ms wide, the average maximal error relative to the range of input signal is 16.84 %. If the measurement interval is extended to 2s, the average maximal error relative to the range of input signal is significantly lowered - down to 1.37 %. Absolute errors are compared with the error limit recommended by Organisation Internationale de Métrologie Légale (OIML) and with the quantization steps of the advanced EEG instruments with 24-bit A/D conversion

Keywords