Applied Sciences (May 2023)

Response Time of a Thin-Film Resistance Temperature Detector (RTD) for High-Intensity Focused Ultrasound (HIFU) Phantom Applications

  • Ma. Stella Grace B. Lequiron,
  • Daniel J. Joe,
  • Sun Do Lim,
  • Il Doh

DOI
https://doi.org/10.3390/app13106220
Journal volume & issue
Vol. 13, no. 10
p. 6220

Abstract

Read online

The thermal response time of a thin-film resistance temperature detector (RTD) array sensor was measured for a high-intensity focused ultrasound (HIFU) phantom. As the temperature inside materials change rapidly within several seconds, it is important to have a temperature sensor with a fast response time to evaluate their performance. However, previous methods for measuring thermal response time were not suitable for thin-film sensors, and there were no quantitative data available. In this study, we used a liquid drop method to measure the thermal time constant of the thin-film RTD, which was found to be 1.0 ± 0.2 ms. This indicates that the thin-film RTD array sensor has a sufficiently fast response time to detect sudden temperature changes inside the tissue-mimicking material (TMM) for validating HIFU devices.