Journal of King Saud University: Science (Oct 2024)

Safety and human health: The landscape of an effective UPLC-MS/MS method for the identification and quantification of N-Nitroso Hydrochlorothiazide impurity in Hydrochlorothiazide

  • Anuradha Bhimireddy,
  • J.V. Shanmukha Kumar,
  • H.N. Prasada Reddy Chittireddy,
  • Mohammed Rafi Shaik,
  • Mohamed E. Assal,
  • Mujeeb Khan

Journal volume & issue
Vol. 36, no. 9
p. 103349

Abstract

Read online

N-nitrosamines (NSAs) and Nitrosamine Drug Substance Related Impurities (NDSRIs) are class of compounds that contains the nitroso functional group. These compounds are carcinogens and mutagens for both humans and animals. As a result, all global regulatory agenesis are requiring risk assessments of NSAs in each pharmaceutical product. The present research aims to provide a single pre-packaged mass spectrometry method parameter set, which includes Ultra performance liquid chromatography and a triple quadrupole mass spectrometer (UPLC-MS-MS) configuration technique for qualifying and quantifying NNHCT in HCT. The analysis was performed on a YMC ODS-A liquid chromatography column with an oven temperature of 35 °C. In this method, mobile phase was prepared by mixing 10 mM ammonium formate buffer and acetonitrile was mixed in the ratio of 60:40 (v/v) with an isocratic program and a flow rate of 0.8 mL/min. The mass detector was used along with electrospray ionization (ESI) Negative mode with multiple reaction monitoring (MRM). The method has been validated in accordance with the current regulatory guidelines and shown the ability to detect NNHCT with a LOQ of 0.01 ppm and LOD of 0.005 ppm.

Keywords