Química Nova (Jul 2024)

DEVICE FOR CALIBRATING ELECTROCHEMICAL DISSOLVED OXYGEN SENSORS

  • Marcus L. Sousa,
  • Isabela A. Duran,
  • Giovana V. Teixeira,
  • André L. B. Formiga

DOI
https://doi.org/10.21577/0100-4042.20250031
Journal volume & issue
Vol. 48, no. 2

Abstract

Read online Read online

Sensor calibration for dissolved gas is a critical and necessary step to ensure consistent sensor response over time and under different conditions. In this work, we describe a 3D printed device that facilitates the calibration of gas sensors. The device effectively maintained gas concentration in the solution for at least 2 h. Analytical curves were constructed in the pH range of 2 to 12, yielding determination coefficients surpassing 0.98. The chosen method proved suitable for the calibration of the sensors. Furthermore, using accurate and reliable reference sensors contributed to obtaining precise and trustworthy measurements.

Keywords