Diabetology & Metabolic Syndrome (Oct 2023)
MiniMed 780G™ advanced hybrid closed-loop system performance in Egyptian patients with type 1 diabetes across different age groups: evidence from real-world users
Abstract
Abstract Background Advanced hybrid closed loop (AHCL) system provides both automated basal rate and correction boluses to keep glycemic values in a target range. Objectives To evaluate the real-world performance of the MiniMed™ 780G system among different age groups of Egyptian patients with type 1diabetes. Methods One-hundred seven AHCL system users aged from 3 to 71 years were enrolled. Data uploaded by patients were aggregated and analyzed. The mean glucose management indicator (GMI), percentage of time spent within glycemic ranges (TIR), time below range (TBR) and time above range (TAR) were determined. Results Six months after initiating Auto Mode, patients spent a mean of 85.31 ± 22.04% of the time in Auto Mode (SmartGuard) and achieved a mean GMI of 6.95 ± 0.58% compared with 7.9 ± 2.1% before AHCL initiation (p 250 mg/dL, TBR < 70 mg/dL and TBR < 54 mg/dL were significantly decreased (p < 0.001). After initiating AHCL, TIR was greater in children and adults compared with adolescents (82.29 ± 7.22% and 83.86 ± 9.24% versus 78.4 ± 7.34%, respectively; p < 0.05). The total daily dose of insulin was increased in all age groups primarily due to increased system-initiated insulin delivery including auto correction boluses and basal insulin. Conclusions MiniMed™ 780G system users across different age groups achieved international consensus-recommended glycemic control with no serious adverse effects even in challenging age group as children and adolescents.