EPJ Web of Conferences (Jan 2016)

Accuracy of Linear Depolarisation Ratios in Clean Air Ranges Measured with POLIS-6 at 355 and 532 NM

  • Freudenthaler Volker,
  • Seefeldner Meinhard,
  • Groß Silke,
  • Wandinger Ulla

DOI
https://doi.org/10.1051/epjconf/201611925013
Journal volume & issue
Vol. 119
p. 25013

Abstract

Read online

Linear depolarization ratios in clean air ranges were measured with POLIS-6 at 355 and 532 nm. The mean deviation from the theoretical values, including the rotational Raman lines within the filter bandwidths, amounts to 0.0005 at 355 nm and to 0.0012 at 532 nm. The mean uncertainty of the measured linear depolarization ratio of clean air is about 0.0005 at 355 nm and about 0.0006 at 532 nm.