GMS Hygiene and Infection Control (May 2021)

Preclinical evaluation of passive disinfection caps with a long-term catheter for the prevention of catheter-related bloodstream infection in pediatric cancer patients

  • Simon, Arne,
  • Fahrendorf, Wolfgang,
  • Hitschmann, Guido

DOI
https://doi.org/10.3205/dgkh000391
Journal volume & issue
Vol. 16
p. Doc20

Abstract

Read online

The use of passive disinfection devices (disinfection caps) may be a beneficial part of a maintenance care bundle, aiming at the prevention of catheter-related bloodstream infections in pediatric cancer patients. This preclinical investigation tested the visual and mechanical integrity of a Broviac™ catheter hub after simulation testing with 122 3M™ CurosTM Stopper Disinfection Caps for Open Female Luers repeatedly attached and removed over 6 months. We found that these catheter hubs were compatible, fully operational, and airtight with use of 3M Curos stopper caps after 6 months of use with 122 caps per catheter hub.