EJNMMI Physics (Jun 2020)

Evaluation of block-sequential regularized expectation maximization reconstruction of 68Ga-DOTATOC, 18F-fluoride, and 11C-acetate whole-body examinations acquired on a digital time-of-flight PET/CT scanner

  • Elin Lindström,
  • Lars Lindsjö,
  • Anders Sundin,
  • Jens Sörensen,
  • Mark Lubberink

DOI
https://doi.org/10.1186/s40658-020-00310-1
Journal volume & issue
Vol. 7, no. 1
pp. 1 – 14

Abstract

Read online

Abstract Background Block-sequential regularized expectation maximization (BSREM) is a fully convergent iterative image reconstruction algorithm. We hypothesize that tracers with different distribution patterns will result in different optimal settings for the BSREM algorithm. The aim of this study was to evaluate the image quality with variations in the applied β-value and acquisition time for three positron emission tomography (PET) tracers. NEMA image quality phantom measurements and clinical whole-body digital time-of-flight (TOF) PET/computed tomography (CT) examinations with 68Ga-DOTATOC (n = 13), 18F-fluoride (n = 10), and 11C-acetate (n = 13) were included. Each scan was reconstructed using BSREM with β-values of 133, 267, 400, and 533, and ordered subsets expectation maximization (OSEM; 3 iterations, 16 subsets, and 5-mm Gaussian post-processing filter). Both reconstruction methods included TOF and point spread function (PSF) recovery. Quantitative measures of noise, signal-to-noise ratio (SNR), and signal-to-background ratio (SBR) were analysed for various acquisition times per bed position (bp). Results The highest β-value resulted in the lowest level of noise, which in turn resulted in the highest SNR and lowest SBR. Noise levels equal to or lower than those of OSEM were found with β-values equal to or higher than 400, 533, and 267 for 68Ga-DOTATOC, 18F-fluoride, and 11C-acetate, respectively. The specified β-ranges resulted in increased SNR at a minimum of 25% (P < 0.0001) and SBR at a maximum of 23% (P < 0.0001) as compared to OSEM. At a reduced acquisition time by 25% for 68Ga-DOTATOC and 18F-fluoride, and 67% for 11C-acetate, BSREM with β-values equal to or higher than 533 resulted in noise equal to or lower than that of OSEM at full acquisition duration (2 min/bp for 68Ga-DOTATOC and 18F-fluoride, 3 min/bp for 11C-acetate). The reduced acquisition time with β 533 resulted in increased SNR (16–26%, P < 0.003) and SBR (6–18%, P < 0.0001 (P = 0.07 for 11C-acetate)) compared to the full acquisition OSEM. Conclusions Within tracer-specific ranges of β-values, BSREM reconstruction resulted in increased SNR and SBR with respect to conventional OSEM reconstruction. Similar SNR, SBR, and noise levels could be attained with BSREM at relatively shorter acquisition times or, alternatively, lower administered dosages, compared to those attained with OSEM.

Keywords