Current Directions in Biomedical Engineering (Sep 2016)

Development of an artifact-free aneurysm clip

  • Brack Alexander,
  • Senger Sebastian,
  • Fischer Gerrit,
  • Janssen Henning,
  • Oertel Joachim,
  • Brecher Christian

DOI
https://doi.org/10.1515/cdbme-2016-0120
Journal volume & issue
Vol. 2, no. 1
pp. 543 – 546

Abstract

Read online

For the treatment of intracranial aneurysms with aneurysm clips, usually a follow-up inspection in MRI is required. To avoid any artifacts, which can make a proper diagnosis difficult, a new approach for the manufacturing of an aneurysm clip entirely made from fiber-reinforced plastics has been developed. In this paper the concept for the design of the clip, the development of a new manufacturing technology for the fiber-reinforced components as well as first results from the examination of the components in phantom MRI testing is shown.

Keywords