Bioengineering (May 2024)

Dislodgment Effects of Different Cage Arrangements in Posterior Lumbar Interbody Fusion: A Finite Element Study

  • Shih-Chieh Yang,
  • Chih-Lin Wu,
  • Yuan-Kun Tu,
  • Pao-Hsin Liu

DOI
https://doi.org/10.3390/bioengineering11060558
Journal volume & issue
Vol. 11, no. 6
p. 558

Abstract

Read online

The vertebral cage has been widely used in posterior lumbar interbody fusion. The risk of cage dislodgment is high for patients undergoing lumbar fusion surgery. Therefore, the main objective of this study was to use a lumbar fusion model to investigate the effects of cage dislodgment on different cage arrangements after PLIF. Finite element analysis was used to compare three PEEK cage placements, together with the fibula-type cage, with respect to the four kinds of lumbar movements. The results revealed that a horizontal cage arrangement could provide a better ability to resist cage dislodgment. Overall lumbar flexion movements were confirmed to produce a greater amount of cage slip than the other three lumbar movements. The lower part of the lumbar fusion segment could create a greater amount of cage dislodgment for all of the lumbar movements. Using an autograft with a fibula as a vertebral cage cannot effectively reduce cage dislodgment. Considering the maximum movement type in lumbar flexion, we suggest that a horizontal arrangement of the PEEK cage might be considered when a single PEEK cage is placed in the fusion segment, as doing so can effectively reduce the extent of cage dislodgment.