Royal Society Open Science (Oct 2020)

Development of a specific troponin I detection system with enhanced immune sensitivity using a single monoclonal antibody

  • Anıl Bozdogan,
  • Reham F. El-Kased,
  • Vanessa Jungbluth,
  • Wolfgang Knoll,
  • Jakub Dostalek,
  • Amal Kasry

DOI
https://doi.org/10.1098/rsos.200871
Journal volume & issue
Vol. 7, no. 10

Abstract

Read online

Using an immunoassay in combination with surface plasmon fluorescence spectroscopy (SPFS), we report the rapid detection of troponin I, a valuable biomarker for diagnosis of myocardial infarction. We discuss the implementation of (i) direct, (ii) sandwich, and (iii) competitive assay formats, based on surface plasmon resonance and SPFS. To elucidate the results, we relate the experiments to orientation-dependent interaction of troponin I epitopes with respective immunoglobulin G antibodies. A limit of detection (LoD) of 19 pM, with 45 min readout time, was achieved using single monoclonal antibody that is specific for one epitope. The borderline between normal people and patients is 20 pM to 83 pM cTnI concentration, and upon the outbreak of acute myocardial infraction it can raise to 2 nM and levels at 20 nM for 6–8 days, therefore the achieved LoD covers most of the clinically relevant range. In addition, this system allows for the detection of troponin I using a single specific monoclonal antibody, which is highly beneficial in case of detection in real samples, where the protein has a complex form leading to hidden epitopes, thus paving the way towards a system that can improve early-stage screening of heart attacks.

Keywords