Journal of Functional Biomaterials (Aug 2024)

Surface Analysis of Orthodontic Mini-Implants after Their Clinical Use

  • Tamara Rahela Ioana,
  • Filip George Boeru,
  • Iulian Antoniac,
  • Ioana Mitruț,
  • Ionela Elisabeta Staicu,
  • Anne Marie Rauten,
  • Willi Andrei Uriciuc,
  • Horia Octavian Manolea

DOI
https://doi.org/10.3390/jfb15090244
Journal volume & issue
Vol. 15, no. 9
p. 244

Abstract

Read online

Temporary anchorage devices (TADs) are orthodontic mini-implants with remarkable characteristics that, once inserted, present mechanical retention (primary stability) without the process of bone osseointegration. However, interaction with the biological environment may cause changes in the morphology of the external surface of dental TADs. In this study, we used 17 TADs made of aluminum–vanadium titanium alloy, produced by two companies, which were analyzed through optical microscopy after being removed from the patients during orthodontic treatment. We evaluated the changes that appeared on the TADs’ surfaces after their use in the biological environment, depending on the morphological area in which they were inserted. In our study, we found changes in the morphology of the implant surface, and especially deposits of biological material in all study groups. On all samples examined after clinical use, regardless of the period of use, corrosion surfaces in different locations were observed. Our obtained results support the idea that the biological environment is aggressive for mini-implant structures, always producing changes to their surface during their clinical use.