EJNMMI Physics (Jan 2024)

The determination of the optimal threshold on measurement of thyroid volume using quantitative SPECT/CT for Graves' hyperthyroidism

  • Chengpeng Gong,
  • Yajing Zhang,
  • Fei Feng,
  • Mengmeng Hu,
  • Kun Li,
  • Rundong Pi,
  • Hua Shu,
  • Rongmei Tang,
  • Xiaoli Wang,
  • Shilin Tan,
  • Fan Hu,
  • Jia Hu

DOI
https://doi.org/10.1186/s40658-023-00608-w
Journal volume & issue
Vol. 11, no. 1
pp. 1 – 13

Abstract

Read online

Abstract Purpose To investigate the optimal threshold for measuring thyroid volume in patients with Grave's hyperthyroidism (GH) by SPECT/CT. Materials and methods A 53 mL butterfly-shaped hollow container made of two 45-degree transparent elbows was put into a NEMA IEC phantom tank. The butterfly-shaped container and the tank were then filled with Na99mTcO4 of different radioactive concentrations, respectively, which could simulate thyroid gland with GH by different target-to-background ratios (T/B) (200:1, 600:1, 1000:1). The different T/B of planar imaging and SPECT/CT were acquired by a Discovery NM/CT 670 Pro SPECT/CT. With Thyroid software (Version 4.0) of GE-Xeleris workstation, the region of the thyroid gland in planar imaging was delineated. The thyroid area and average long diameter of both lobes were substituted into the Allen formula to calculate the thyroid volume. The calculation error was compared with the actual volume. Q-Metrix software was used to perform CT-based attenuation correction, scatter correction, resolution recovery. Ordered-subsets expectation maximization was used to reconstruct SPECT data. 20%, 25%, 30%, 40%, 50%, 60% thresholds were selected to automatically delineate the volume of interest and compared with the real volume, which determinated the optimal threshold. We measured the thyroid volume of 40 GH patients using the threshold and compared the volumes obtained by planar imaging and ultrasound three-dimensional. The differences of the volumes with different T/B and thresholds were compared by the ANOVA and least significant difference t test. The volumes delineated by SPECT/CT were evaluated using ANOVA, least significant difference t test, correlation analysis and, linear regression and Bland–Altman concordance test plot. The differences and consistency of thyroid volume were compared among the above three methods. Results There was no significant difference in the results between different T/B models (P > 0.05). The thyroid volume calculated by the planar imaging formula method was higher than the real volume, with an average overestimation of 22.81%. The volumes delineated by SPECT/CT threshold automatically decreased while the threshold increased. There were significant differences between groups with different thresholds (P 0.05). Thyroid volume measured by planar imaging method was significantly higher than ultrasound and SPECT/CT threshold automatic delineation method (P < 0.05). The agreement between the SPECT/CT 25% threshold and ultrasound (r = 0.956, b = 0.961) was better than that between the planar imaging and ultrasound (r = 0.590, b = 0.574). The Bland–Altman plot also showed that the thyroid volume measured by the 25% threshold automatic delineation method was in good agreement with the ultrasound measurement. Conclusions The T/B has no effect on the measurement of thyroid volume in GH patients; planar imaging method can significantly overestimate thyroid volume in GH patients, and 25% threshold automatic delineation method can obtain more accurate thyroid volume in GH patients.

Keywords