Gynecology and Minimally Invasive Therapy (Aug 2015)

Invention of a new Lin soft outer sheath as a continuous flow system for diagnostic flexible hysteroscopy

  • Bao-Liang Lin,
  • Hsuan Chin,
  • Midori Ookouchi,
  • Takashi Takeda,
  • Takeshi Suzuki,
  • Kazunori Ueno,
  • Sokichi Iwata

DOI
https://doi.org/10.1016/j.gmit.2015.01.007
Journal volume & issue
Vol. 4, no. 3
pp. 87 – 90

Abstract

Read online

Objectives: This study aims to evaluate the efficacy of a newly developed Lin soft outer sheath that is used as a continuous flow system for outpatient flexible hysteroscopy. Materials and methods: Diagnostic hysteroscopy was performed in 134 patients who presented to our center with uterine bleeding over the 25-month study period. The flexible hysteroscope was equipped with a Lin soft outer sheath. Results: Hysteroscope insertion failure was observed in six cases of, i.e., an insertion failure rate of 4.5%. Out of the other 128 women, accurate diagnosis of four patients could not be made because of intrauterine blood clots, i.e., a success rate of 92.5%. No complication, except postprocedure bleeding for a few days, was encountered. Conclusion: The new Lin soft outer sheath is effective as a continuous flow system for outpatient flexible hysteroscopy.

Keywords