Journal of Personalized Medicine (Feb 2023)

Applying Multi-Metric Deformable Image Registration for Dose Accumulation in Combined Cervical Cancer Radiotherapy

  • Qi Fu,
  • Xin Xie,
  • Yingjie Xu,
  • Jing Zuo,
  • Xi Yang,
  • Wenlong Xia,
  • Jusheng An,
  • Manni Huang,
  • Hui Yan,
  • Jianrong Dai

DOI
https://doi.org/10.3390/jpm13020323
Journal volume & issue
Vol. 13, no. 2
p. 323

Abstract

Read online

(1) Purpose: Challenges remain in dose accumulation for cervical cancer radiotherapy combined with external beam radiotherapy (EBRT) and brachytherapy (BT) as there are many large and complex organ deformations between different treatments. This study aims to improve deformable image registration (DIR) accuracy with the introduction of multi-metric objectives for dose accumulation of EBRT and BT. (2) Materials and methods: Twenty cervical cancer patients treated with EBRT (45–50 Gy/25 fractions) and high-dose-rate BT (≥20 Gy in 4 fractions) were included for DIR. The multi-metric DIR algorithm included an intensity-based metric, three contour-based metrics, and a penalty term. Nonrigid B-spine transformation was used to transform the planning CT images from EBRT to the first BT, with a six-level resolution registration strategy. To evaluate its performance, the multi-metric DIR was compared with a hybrid DIR provided by commercial software. The DIR accuracy was measured by the Dice similarity coefficient (DSC) and Hausdorff distance (HD) between deformed and reference organ contours. The accumulated maximum dose of 2 cc (D2cc) of the bladder and rectum was calculated and compared to simply addition of D2cc from EBRT and BT (ΔD2cc). (3) Results: The mean DSC of all organ contours for the multi-metric DIR were significantly higher than those for the hybrid DIR (p ≤ 0.011). In total, 70% of patients had DSC > 0.8 using the multi-metric DIR, while 15% of patients had DSC > 0.8 using the commercial hybrid DIR. The mean ΔD2cc of the bladder and rectum for the multi-metric DIR were 3.25 ± 2.29 and 3.54 ± 2.02 GyEQD2, respectively, whereas those for the hybrid DIR were 2.68 ± 2.56 and 2.32 ± 3.25 GyEQD2, respectively. The multi-metric DIR resulted in a much lower proportion of unrealistic D2cc than the hybrid DIR (2.5% vs. 17.5%). (4) Conclusions: Compared with the commercial hybrid DIR, the introduced multi-metric DIR significantly improved the registration accuracy and resulted in a more reasonable accumulated dose distribution.

Keywords