A D-Shaped Bileaflet Bioprosthesis which Replicates Physiological Left Ventricular Flow Patterns.

  • Sean Guo-Dong Tan,
  • Sangho Kim,
  • Jimmy Kim Fatt Hon,
  • Hwa Liang Leo

DOI
https://doi.org/10.1371/journal.pone.0156580
Journal volume & issue
Vol. 11, no. 6
p. e0156580

Abstract

Read online

Prior studies have shown that in a healthy heart, there exist a large asymmetric vortex structure that aids in establishing a steady flow field in the left ventricle. However, the implantation of existing artificial heart valves at the mitral position is found to have a negative effect on this physiological flow pattern. In light of this, a novel D-shaped bileaflet porcine bioprosthesis (GD valve) has been designed based on the native geometry mitral valve, with the hypothesis that biomimicry in valve design can restore physiological left ventricle flow patterns after valve implantation. An in-vitro experiment using two dimensional particle velocimetry imaging was carried out to determine the hemodynamic performance of the new bileaflet design and then compared to that of the well-established St. Jude Epic valve which functioned as a control in the experiment. Although both valves were found to have similar Reynolds shear stress and Turbulent Kinetic Energy levels, the novel D-shape valve was found to have lower turbulence intensity and greater mean kinetic energy conservation.