Open Medicine (Feb 2022)
CRRT influences PICCO measurements in febrile critically ill patients
Abstract
The aim of this study was to investigate whether continuous renal replacement therapy (CRRT) influences the global end-diastolic volume index (GEDVI), cardiac index (CI), and extravascular lung water index (EVLWI) measured by Pulse Index Continuous Cardiac Output (PICCO) in febrile patients. Fifteen fever patients were included in this study. CI, GEDVI, EVLWI, heart rate (HR), and mean arterial pressure (MAP) were measured at five time-points: before CRRT (T0), immediately after CRRT started (T1), 15 min after CRRT started (T2), immediately after CRRT stopped (T3), and 15 min after CRRT stopped (T4). Results have shown that CI and GEDVI were decreased significantly in T1 (CI: 4.09 ± 0.72 vs 2.81 ± 0.58 L/min m2, P = 0.000 and GEDVI: 727.86 ± 63.47 vs 531.07 ± 66.63 mL/m2, P = 0.000). However, CI and GEDVI were significantly increased in T3 (CI: 4.09 ± 0.72 vs 7.23 ± 1.32 L/min m2, P = 0.000 and GEDVI 727.86 ± 63.47 vs 1339.17 ± 121.52 mL/m2, P = 0.000). There were no significant differences in T2 and T4. Among the five-time points, no measurement errors were observed with regards to HR, MAP, and EVLWI. Therefore, the data herein contained suggests that PICCO measurements should begin 15 min after the start or stop of CRRT.
Keywords