Brazilian Journal of Radiation Sciences (Apr 2021)

Comparative evaluation of the performance of thin diodes used as on-line dosimeters in radiation processing applications

  • Josemary A. C. Gonçalves,
  • Alessio Mangiarotti,
  • Carmen C Bueno

DOI
https://doi.org/10.15392/bjrs.v9i1A.1327
Journal volume & issue
Vol. 9, no. 1A

Abstract

Read online

In this work, we report a comparison between the performance of two samples of commercial PIN photodiodes (SFH206K from Osram and S2506-04 from Hamamatsu) mainly addressing the variation of their current sensitivities with accumulated dose ranging from 0-15 kGy. All the results so far obtained have revealed that the radiation induced currents are linearly dependent on dose rates from 3.65 to 55.64 Gy/h. The current sensitivity of both unirradiated diodes (0.178 nA.h/Gy.mm3) slightly decreases with accumulated dose, namely 0.32%/kGy (SFH206K) and 1.4%/kGy (S2506-04). Although the SFH206K device compares favorably with the S2506-04, both diodes can be considered as a low budge alternative, good enough for on-line dosimetry applications in the field of radiation processing.