Canadian Respiratory Journal (Jan 2017)

Insufficient Humidification of Respiratory Gases in Patients Who Are Undergoing Therapeutic Hypothermia at a Paediatric and Adult Intensive Care Unit

  • Yukari Tanaka,
  • Sachiko Iwata,
  • Masahiro Kinoshita,
  • Kennosuke Tsuda,
  • Shoichiro Tanaka,
  • Naoko Hara,
  • Ryota Shindou,
  • Eimei Harada,
  • Ryouji Kijima,
  • Osamu Yamaga,
  • Hitoe Ohkuma,
  • Kazuo Ushijima,
  • Teruo Sakamoto,
  • Yushiro Yamashita,
  • Osuke Iwata

DOI
https://doi.org/10.1155/2017/8349874
Journal volume & issue
Vol. 2017

Abstract

Read online

For cooled newborn infants, humidifier settings for normothermic condition provide excessive gas humidity because absolute humidity at saturation is temperature-dependent. To assess humidification of respiratory gases in patients who underwent moderate therapeutic hypothermia at a paediatric/adult intensive care unit, 6 patients were studied over 9 times. Three humidifier settings, 37-default (chamber-outlet, 37°C; Y-piece, 40°C), 33.5-theoretical (chamber-outlet, 33.5°C; Y-piece, 36.5°C), and 33.5-adjusted (optimised setting to achieve saturated vapour at 33.5°C using feedback from a thermohygrometer), were tested. Y-piece gas temperature/humidity and the incidence of high (>40.6 mg/L) and low (<32.9 mg/L) humidity relative to the target level (36.6 mg/L) were assessed. Y-piece gas humidity was 32.0 (26.8–37.3), 22.7 (16.9–28.6), and 36.9 (35.5–38.3) mg/L {mean (95% confidence interval)} for 37-default setting, 33.5-theoretical setting, and 33.5-adjusted setting, respectively. High humidity was observed in 1 patient with 37-default setting, whereas low humidity was seen in 5 patients with 37-default setting and 8 patients with 33.5-theoretical setting. With 33.5-adjusted setting, inadequate Y-piece humidity was not observed. Potential risks of the default humidifier setting for insufficient respiratory gas humidification were highlighted in patients cooled at a paediatric/adult intensive care unit. Y-piece gas conditions can be controlled to the theoretically optimal level by adjusting the setting guided by Y-piece gas temperature/humidity.