Current Directions in Biomedical Engineering (Sep 2016)

Contactless respiratory monitoring system for magnetic resonance imaging applications using a laser range sensor

  • Krug Johannes W.,
  • Odenbach Robert,
  • Boese Axel,
  • Friebe Michael

DOI
https://doi.org/10.1515/cdbme-2016-0156
Journal volume & issue
Vol. 2, no. 1
pp. 719 – 722

Abstract

Read online

During a magnetic resonance imaging (MRI) exam, a respiratory signal can be required for different purposes, e.g. for patient monitoring, motion compensation or for research studies such as in functional MRI. In addition, respiratory information can be used as a biofeedback for the patient in order to control breath holds or shallow breathing. To reduce patient preparation time or distortions of the MR imaging system, we propose the use of a contactless approach for gathering information about the respiratory activity. An experimental setup based on a commercially available laser range sensor was used to detect respiratory induced motion of the chest or abdomen. This setup was tested using a motion phantom and different human subjects in an MRI scanner. A nasal airflow sensor served as a reference. For both, the phantom as well as the different human subjects, the motion frequency was precisely measured. These results show that a low cost, contactless, laser-based approach can be used to obtain information about the respiratory motion during an MRI exam.

Keywords