A Clinical Infrared Video-Oculoscopy Suppression Head Impulse (IR-cSHIMP) Test

  • Vincenzo Marcelli,
  • Beatrice Giannoni

DOI
https://doi.org/10.3390/audiolres14010013
Journal volume & issue
Vol. 14, no. 1
pp. 151 – 165

Abstract

Read online

Background: We propose a Suppression Head IMPulse (SHIMP) test method that provides for equipment only through the use of InfraRed Video-OculoScopy (IR-VOS) and allows horizontal and vertical semicircular canal function evaluation in bedside mode. We therefore named the test InfraRed clinical SHIMP (IR-cSHIMP). Methods: To check IR-cSHIMP diagnostic efficiency, we studied 22 normal subjects, 18 patients with unilateral, and 6 with bilateral deficient vestibulopathy. Each subject first underwent a vestibular examination and, only later, an IRc-SHIMP test. Results: When the IR-cSHIMP test was performed in the horizontal plane, all normal subjects showed anti-compensatory saccades. When the vertical semicircular canal function was evaluated, the same result was obtained in all normal subjects except three, which were considered false positives. In patients with vestibular deficits, the test performed in the horizontal and vertical planes were always pathological, with 100% agreement between clinical and instrumental tests. Conclusions: Our bedside method proved to be fast, simple, and effective in discriminating between healthy and pathological subjects. It required only the same skill as the better-known cHIT. For these reasons, we believe that the IR-cSHIMP should be part of daily clinical practice as a useful tool in the selection of patients to undergo more sophisticated investigations.

Keywords