Physics and Imaging in Radiation Oncology (Jul 2022)

Linear energy transfer-independent calibration of radiochromic film for carbon-ion beams

  • Mutsumi Tashiro,
  • Motohiro Kawashima

Journal volume & issue
Vol. 23
pp. 140 – 143

Abstract

Read online

For carbon-ion beams, radiochromic film response depends on the dose and linear energy transfer (LET). For film dosimetry, we developed an LET-independent simple calibration method for a radiochromic film for specific therapeutic carbon-ion beams. The measured film doses were calibrated with a linear function within 5% error. The penumbra positions of the films were consistent with the differences from the planned ones within ~0.4 mm. The results indicated sufficient accuracy for use as a tool for the confirmation of the penumbra position of the fields.