Communications Medicine (Jul 2024)
Recommendations for evaluating photoplethysmography-based algorithms for blood pressure assessment
Abstract
Abstract Photoplethysmography (PPG) is a non-invasive optical technique that measures changes in blood volume in the microvascular tissue bed of the body. While it shows potential as a clinical tool for blood pressure (BP) assessment and hypertension management, several sources of error can affect its performance. One such source is the PPG-based algorithm, which can lead to measurement bias and inaccuracy. Here, we review seven widely used measures to assess PPG-based algorithm performance and recommend implementing standardized error evaluation steps in their development. This standardization can reduce bias and improve the reliability and accuracy of PPG-based BP estimation, leading to better health outcomes for patients managing hypertension.