Патология кровообращения и кардиохирургия (Jun 2018)

Aortic valve replacement with stentless bioprosthesis «Kemerovo-AB-Neo»

  • D. A. Astapov,
  • A. M. Karaskov,
  • Ye. I. Semenova,
  • D. P. Demidov,
  • M. V. Isayan,
  • A. B. Open

DOI
https://doi.org/10.21688/1681-3472-2012-1-23-26
Journal volume & issue
Vol. 16, no. 1
pp. 23 – 26

Abstract

Read online

We analyzed 50 aortic valve replacements (AVR) with Kemerovo-AB-Neo stentless bioprostheses. Mean age was 66.75 (54 to78) years, hospital mortality rate accounted for 2%. The peak transprosthetic pressure gradient (PTPG) in patients operated for aortic stenosis came to 19.57 mm Hg. PTPG was shown to be dependent on the implantation technique; it reached 24.57 mm Hg when in order to fix the proximal line, interrupted sutures were used and ran to 175 mm Hg in the case of running sutures (р = 0.03). Helical CT confirmed fine mobility of the cardiac structures after Kemerovo-AB-Neo implantation: the aortic cross-section area varied up to 84% during the cardiac cycle. It should be noted that AVR with Kemerovo-AB-Neo stentless valves gives good clinical and hemodynamic results early after surgery. A free-hand technique of implantation should be preferred. The implantation of stentless bioprostheses retains cardiac structures mobility and natural aortic root dimensions after performing an AVR.