Chronoamperometric Ammonium Ion Detection in Water via Conductive Polymers and Gold Nanoparticles
Roberta Farina,
Silvia Scalese,
Domenico Corso,
Giuseppe Emanuele Capuano,
Giuseppe Andrea Screpis,
Maria Anna Coniglio,
Guglielmo Guido Condorelli,
Sebania Libertino
Affiliations
Roberta Farina
Consiglio Nazionale delle Ricerche-Istituto per la Microelettronica e Microsistemi (CNR-IMM), Strada VIII Z.I., 5, 95121 Catania, Italy
Silvia Scalese
Consiglio Nazionale delle Ricerche-Istituto per la Microelettronica e Microsistemi (CNR-IMM), Strada VIII Z.I., 5, 95121 Catania, Italy
Domenico Corso
Consiglio Nazionale delle Ricerche-Istituto per la Microelettronica e Microsistemi (CNR-IMM), Strada VIII Z.I., 5, 95121 Catania, Italy
Giuseppe Emanuele Capuano
Consiglio Nazionale delle Ricerche-Istituto per la Microelettronica e Microsistemi (CNR-IMM), Strada VIII Z.I., 5, 95121 Catania, Italy
Giuseppe Andrea Screpis
Dipartimento di Scienze Mediche, Chirurgiche e Tecnologie Avanzate “G.F. Ingrassia”, Università Degli Studi di Catania, Via S. Sofia 87, 95123 Catania, Italy
Maria Anna Coniglio
Consiglio Nazionale delle Ricerche-Istituto per la Microelettronica e Microsistemi (CNR-IMM), Strada VIII Z.I., 5, 95121 Catania, Italy
Guglielmo Guido Condorelli
Dipartimento di Scienze Chimiche, Università Degli Studi di Catania, Viale A. Doria 6, 95125 Catania, Italy
Sebania Libertino
Consiglio Nazionale delle Ricerche-Istituto per la Microelettronica e Microsistemi (CNR-IMM), Strada VIII Z.I., 5, 95121 Catania, Italy
Monitoring of ammonium ion levels in water is essential due to its significant impact on environmental and human health. This work aims to fabricate and characterize sensitive, real-time, low-cost, and portable amperometric sensors for low NH4+ concentrations in water. Two strategies were conducted by cyclic voltammetry (CV): electrodeposition of Au nanoparticles on a commercial polyaniline/C electrode (Au/PANI/C), and CV of electropolymerized polyaniline on a commercial carbon electrode (Au/PANIep/C). Au NPs increase the electrical conductivity of PANI and its ability to transfer charges during electrochemical reactions. The electrode performances were tested in a concentration range from 0.35 µM to 7 µM in NH4+ solution. The results show that the Au/PANI/C electrode performs well for high NH4+ concentrations (0.34 µM LoD) and worsens for low NH4+ concentrations (0.01 µM LoD). A reverse performance occurs for the electrode Au/PANIep/C, with a 0.03 µM LoD at low NH4+ concentration and 0.07 µM LoD at high NH4+ concentration. The electrodes exhibit a good reproducibility, with a maximum RSD of 3.68% for Au/PANI/C and 5.94% for Au/PANIep/C. In addition, the results of the repeatability tests show that the electrochemical reaction of sensing is fully reversible, leaving the electrode ready for a new detection event.