PLoS ONE (Jan 2015)

Compass: clinical evaluation of a new instrument for the diagnosis of glaucoma.

  • Luca Rossetti,
  • Maurizio Digiuni,
  • Alberto Rosso,
  • Roberta Riva,
  • Giuliano Barbaro,
  • Michael K Smolek,
  • Nicola Orzalesi,
  • Stefano De Cilla',
  • Alessandro Autelitano,
  • Paolo Fogagnolo

DOI
https://doi.org/10.1371/journal.pone.0122157
Journal volume & issue
Vol. 10, no. 3
p. e0122157

Abstract

Read online

To evaluate Compass, a new instrument for glaucoma screening and diagnosis that combines scanning ophthalmoscopy, automated perimetry, and eye tracking.A total of 320 human subjects (200 normal, 120 with glaucoma) underwent full ophthalmological evaluation and perimetric evaluation using the Humphrey SITA standard 24° test (HFA), and the Compass test that consisted of a full-threshold program on the central 24° with a photograph of the central 30° of the retina. A subgroup of normal subjects and glaucoma patients underwent a second Compass test during the same day in order to study test-retest variability. After exclusion of 30 patients due to protocol rules, a database was created to compare the Compass to the HFA, and to evaluate retinal image quality and fixation stability.The difference in mean sensitivity between Compass and HFA was -1.02 ± 1.55 dB in normal subjects (p65% of cases; Image-based diagnosis was in accordance with the initial diagnosis in 85% of the subjects.Based on preliminary results, Compass showed useful diagnostic characteristics for the study of glaucoma, and combined morphological information with functional data.