Journal of Sensors and Sensor Systems (Nov 2018)

Precise linear measurements using a calibrated reference workpiece without temperature measurements

  • D. Sumin,
  • R. Tutsch

DOI
https://doi.org/10.5194/jsss-7-609-2018
Journal volume & issue
Vol. 7
pp. 609 – 620

Abstract

Read online

We suggest a procedure for the correction of the errors caused by thermal expansion of a workpiece and the scale of a linear measuring instrument (coordinate measuring machines, length measuring machines, etc.) when linear measurements are performed at nonstandard temperature. We use a calibrated reference workpiece but do not require temperature measurements. An estimation of the measurement uncertainty and application examples are given.