Sensors (Aug 2022)

Silicon Cantilever for Micro/Nanoforce and Stiffness Calibration

  • Joachim Frühauf,
  • Eva Gärtner,
  • Zhi Li,
  • Lutz Doering,
  • Jan Spichtinger,
  • Gerd Ehret

DOI
https://doi.org/10.3390/s22166253
Journal volume & issue
Vol. 22, no. 16
p. 6253

Abstract

Read online

The paper deals with cantilevers made from monocrystalline silicon by processes of microtechnology. The cantilevers are passive structures and have no transducers. The application as a material measure for the inspection of stylus forces is in the center of investigations. A simple method is the measurement of the deflection of the cantilever at the position of load by the force if the stiffness of the cantilever at this position is known. Measurements of force–deflection characteristics are described and discussed in context with the classical theory of elastic bending. The methods of determining the stiffness are discussed together with results. Finally, other methods based on tactile measurements along the cantilever are described and tested. The paper discusses comprehensively the properties of concrete silicon chips with cantilevers to underpin its applicability in industrial metrology. The progress consists of the estimation of the accuracy of the proposed method of stylus force measurement and the extraction of information from a tactile measured profile along the silicon cantilever. Furthermore, improvements are proposed for approaches to an ideal cantilever.

Keywords