International Journal of Ophthalmology (Jan 2022)
Twenty-four hours intraocular pressure in keratoconic eyes assessed by applanation tonometry and Tono-Pen AVIA
Abstract
AIM: To assess intraocular pressure (IOP) during the daily curve of intraocular pressure (DCPo) in keratoconic eyes and compare Goldmann applanation tonometer (GAT), without and with astigmatism correction (nGAT and cGAT) and Tono-Pen AVIA (TPA) assessment methods. METHODS: Thirty-nine keratoconic eyes of 24 patients were assessed. DCPo was evaluated with five IOP measurements; four were performed with a GAT (nGAT and cGAT), and a Tono-Pen AVIA (TPA) at various times throughout the day. RESULTS: Mean IOP DCPo values (mm Hg) were: nGAT, 9.9±2.6; cGAT, 11.3±2.6; TPA 12.3±3.1. Mean IOP DCPo differences (mm Hg) and Spearman's correlation coefficients were as follows: cGATc-nGAT, 1.32±1.31, rs=0.879 (PcGAT>nGAT. IOP TPA measures are predictive of cGAT values, adjusted according to anterior chamber depth and corneal astigmatism.
Keywords