Retrospectively ECG-gated helical vs. non-ECG-synchronized high-pitch CTA of the aortic root for TAVI planning.

  • Barbora Horehledova,
  • Casper Mihl,
  • Ellen Boswijk,
  • Genevieve A J C Crombag,
  • Estelle C Nijssen,
  • Patty J Nelemans,
  • Leo F Veenstra,
  • Joachim E Wildberger,
  • Marco Das

DOI
https://doi.org/10.1371/journal.pone.0232673
Journal volume & issue
Vol. 15, no. 5
p. e0232673

Abstract

Read online

BACKGROUND:Multidetector computed tomography (MDCT) plays a key role in patient assessment prior to transcatheter aortic valve implantation (TAVI). However, to date no consensus has been established on what is the optimal pre-procedural imaging protocol. Variability in pre-TAVI acquisition protocols may lead to discrepancies in aortic annulus measurements and may potentially influence prosthesis size selection. PURPOSE:The current study evaluates the magnitude of differences in aortic annulus measurements using max-systolic, end-diastolic, and non-ECG-synchronized imaging, as well as the impact of method on prosthesis size selection. MATERIAL AND METHODS:Fifty consecutive TAVI-candidates, who underwent retrospectively-ECG-gated CT angiography (CTA) of the aortic root, directly followed by non-ECG-synchronized high-pitch CT of the entire aorta, were retrospectively included. Aortic root dimensions were assessed at each 10% increment of the R-R interval (0-100%) and on the non-ECG-synchronized scan. Dimensional changes within the cardiac cycle were evaluated using a 1-way repeated ANOVA. Agreement in measurements between max-systole, end-diastole and non-ECG-synchronized scans was assessed with Bland-Altman analysis. RESULTS:Maximal dimensions of the aortic root structures and minimum annulus-coronary ostia distances were measured during systole. Max-systolic measurements were significantly and substantially larger than end-diastolic (p<0.001) and non-ECG-synchronized measurements (p<0.001). Due to these discrepancies, the three methods resulted in the same prosthesis size selection in only 48-62% of patients. CONCLUSIONS:The systematic differences between max-systolic, end-diastolic and non-ECG-synchronized measurements for relevant aortic annular dimensions are both statistically significant and clinically relevant. Imaging strategy impacts prosthesis size selection in nearly half the TAVI-candidates. End-diastolic and non-ECG-synchronized imaging does not provide optimal information for prosthesis size selection. Systolic image acquisition is necessary for assessment of maximal annular dimensions and minimum annulus-coronary ostia distances.