European Journal of Translational Myology (Jun 2016)

Inductive powering of subcutaneous stimulators: key parameters and their impact on the design methodology

  • Carmen Godfraind,
  • Adrien Debelle,
  • Laurent Lonys,
  • Vicente Acuña,
  • Pascal Doguet,
  • Antoine Nonclercq

DOI
https://doi.org/10.4081/ejtm.2016.6040
Journal volume & issue
Vol. 26, no. 2

Abstract

Read online

Inductive powering of implantable medical devices involves numerous factors acting on the system efficiency and safety in adversarial ways. This paper lightens up their role and identifies a procedure enabling the system design. The latter enables the problem to be decoupled into four principal steps: the frequency choice, the magnetic link optimization, the secondary circuit and then finally the primary circuit designs. The methodology has been tested for the powering system of a device requirering a power of 300mW and implanted at a distance of 15 to 30mm from the outside power source. It allowed the identification of the most critical parameters. A satisfying efficiency of 34% was reached at 21mm and tend to validate the proposed design procedure.