Asian Journal of Medical Sciences (Aug 2022)

Effectiveness of orthopedic metal artifact reduction among patients undergoing computed tomography at a tertiary setting – A cross-sectional study

  • Ramesh Parthasarathy ,
  • Ealai Athmarao Parthasarathy ,
  • Anand Rajamani ,
  • Gurubharath Ilangovan ,
  • Niranjana AK

DOI
https://doi.org/10.3126/ajms.v13i8.44770
Journal volume & issue
Vol. 13, no. 8
pp. 154 – 160

Abstract

Read online

Background: In computed tomography (CT), metal implants increase the inconsistencies between the measured data and the linear assumption of the radon transform made by the analytic CT reconstruction algorithm. Aims and Objectives: The aims of this study was to evaluate the effectiveness of orthopedic metal artifact reduction (O-MAR) among patients undergoing CT at a tertiary setting. Materials and Methods: A cross-sectional observational study was conducted on 30 subjects attending the department of radiology for CT scan for 1 year from December 2018 to 2019. The CT scan was performed in 64 detector row (PHILIPS INGENUITY CORE 128 slice) scanner. The patient was asked for the presence of metallic objects/history of implants, and the topo gram confirms it. Image noise was considered as the primary outcome variable. The statistical analysis was performed using the coGuide software and P<0.005. Results: All the thirty participants were analyzed finally. Image noise in plain image type was 79.5 SD of Hounsfield units, and in O-MAR image type, it was 44.01 SD of Hounsfield units denoting a higher percentage of SD in plain than O-MAR images. Conclusion: The O-MAR application helped in improving image quality of CT for patients in the aspect of metal artifact reduction and soft-tissue profile. However, it can also improve diagnostic quality in the evaluation of patients with severe metallic artifacts by decreasing the negative impacts of orthopedic metals.

Keywords