The International Archives of the Photogrammetry, Remote Sensing and Spatial Information Sciences (Jun 2016)

OBSERVATIONS ON THE PERFORMANCE OF X-RAY COMPUTED TOMOGRAPHY FOR DIMENSIONAL METROLOGY

  • H. C. Corcoran,
  • H. C. Corcoran,
  • S. B. Brown,
  • S. Robson,
  • R. D. Speller,
  • M. B. McCarthy

DOI
https://doi.org/10.5194/isprs-archives-XLI-B5-25-2016
Journal volume & issue
Vol. XLI-B5
pp. 25 – 31

Abstract

Read online

X-ray computed tomography (XCT) is a rising technology within many industries and sectors with a demand for dimensional metrology, defect, void analysis and reverse engineering. There are many variables that can affect the dimensional metrology of objects imaged using XCT, this paper focusses on the effects of beam hardening due to the orientation of the workpiece, in this case a holeplate, and the volume of material the X-rays travel through. Measurements discussed include unidirectional and bidirectional dimensions, radii of cylinders, fit point deviations of the fitted shapes and cylindricity. Results indicate that accuracy and precision of these dimensional measurements are affected in varying amounts, both by the amount of material the X-rays have travelled through and the orientation of the object.