Physics and Imaging in Radiation Oncology (Apr 2018)

A multi-centre analytical study of small field output factor calculations in radiotherapy

  • Krzysztof Chełmiński,
  • Wojciech Bulski

Journal volume & issue
Vol. 6
pp. 1 – 4

Abstract

Read online

An audit methodology was developed and applied for output factor (OF) calculations in radiotherapy. The auditees were asked to calculate OFs for field sizes from 10 × 10 cm2 to 2 × 2 cm2. Sixty five beams were audited; missing reference OFs were interpolated. The calculated OFs were in 73% of cases higher than the reference data. The smaller the field size, the higher the overestimations which were observed in the higher fraction of cases. Treatment planning systems generally overestimated OFs for small fields. The reference dataset helped radiotherapy centres to identify discrepancies which were higher than typical. Keywords: Computer-assisted radiotherapy planning, Intensity-modulated radiotherapy, Particle accelerators, Dosimetry audit, Small field dosimetry