Pharmacia (Jan 2022)

HPLC determination of Escitalopram in tablet dosage forms

  • Stefan Balkanski

DOI
https://doi.org/10.3897/pharmacia.69.e77878
Journal volume & issue
Vol. 69, no. 1
pp. 21 – 24

Abstract

Read online Read online Read online

Purpose: A simple, specific, precise, and accurate reversed phase liquid chromatographic (RP-LC) method has been developed for the determination of Escitalopram in tablet dosage form. Methods: The chromatographic separation was achieved on a LiChrosorb C18, 250 mm x 4.6 mm, 5 μm column at a detector wavelength of 270 nm and a flow rate of 1.0 ml/min. The mobile phase was composed of methanol, acetonitrile (70:30 v/v). The retention time of Escitalopram was 5.49 min. The method was validated for the parameters like specificity, linearity, precision, accuracy, limit of quantitation and limit of detection. Results: The method was found to be specific as no other peaks of impurities and excipients were observed. The square of correlation coefficient (R2) was 0.9999 while relative standard deviations were found to be <2.0%. Conclusion: The proposed RP-LC method can be applied for the routine analysis of commercially available formulations of Escitalopram.