BMC Cancer (Apr 2021)
Analytical validation of the Percepta genomic sequencing classifier; an RNA next generation sequencing assay for the assessment of Lung Cancer risk of suspicious pulmonary nodules
Abstract
Abstract Background Bronchoscopy is a common procedure used for evaluation of suspicious lung nodules, but the low diagnostic sensitivity of bronchoscopy often results in inconclusive results and delays in treatment. Percepta Genomic Sequencing Classifier (GSC) was developed to assist with patient management in cases where bronchoscopy is inconclusive. Studies have shown that exposure to tobacco smoke alters gene expression in airway epithelial cells in a way that indicates an increased risk of developing lung cancer. Percepta GSC leverages this idea of a molecular “field of injury” from smoking and was developed using RNA sequencing data generated from lung bronchial brushings of the upper airway. A Percepta GSC score is calculated from an ensemble of machine learning algorithms utilizing clinical and genomic features and is used to refine a patient’s risk stratification. Methods The objective of the analysis described and reported here is to validate the analytical performance of Percepta GSC. Analytical performance studies characterized the sensitivity of Percepta GSC test results to input RNA quantity, the potentially interfering agents of blood and genomic DNA, and the reproducibility of test results within and between processing runs and between laboratories. Results Varying the amount of input RNA into the assay across a nominal range had no significant impact on Percepta GSC classifier results. Bronchial brushing RNA contaminated with up to 10% genomic DNA by nucleic acid mass also showed no significant difference on classifier results. The addition of blood RNA, a potential contaminant in the bronchial brushing sample, caused no change to classifier results at up to 11% contamination by RNA proportion. Percepta GSC scores were reproducible between runs, within runs, and between laboratories, varying within less than 4% of the total score range (standard deviation of 0.169 for scores on 4.57 scale). Conclusions The analytical sensitivity, analytical specificity, and reproducibility of Percepta GSC laboratory results were successfully demonstrated under conditions of expected day to day variation in testing. Percepta GSC test results are analytically robust and suitable for routine clinical use.
Keywords