Química Nova (Jan 2012)
Controle da qualidade na determinação potenciométrica de sulfetos volatilizáveis por acidificação (SVA) utilizando adições de padrão
Abstract
This paper presents a quality assurance and quality control protocol in order to ensure the reliability of analytical results for the determination of acid volatile sulfides in environmental samples by potentiometry. The monitoring of the ion selective sensor readings must be realized continually and particularly in this situation where the analyte is susceptible to transformation and volatilization. Potentiometry with the standard addition was presented as a rapid, accurate, precise, feasible as well as a low cost technique. Concentrations of the analyte in the range of 0.49 to 1.02 mg L-1 were tested with percentage recoveries varying from 89 to 113%.
Keywords