Scientific Reports (May 2022)
Contrast timing optimization of a two-volume dynamic CT pulmonary perfusion technique
Abstract
Abstract The purpose of this study is to develop and validate an optimal timing protocol for a low-radiation-dose CT pulmonary perfusion technique using only two volume scans. A total of 24 swine (48.5 ± 14.3 kg) underwent contrast-enhanced dynamic CT. Multiple contrast injections were made under different pulmonary perfusion conditions, resulting in a total of 141 complete pulmonary arterial input functions (AIFs). Using all the AIF curves, an optimal contrast timing protocol was developed for a first-pass, two-volume dynamic CT perfusion technique (one at the base and the other at the peak of AIF curve). A subset of swine was used to validate the prospective two-volume pulmonary perfusion technique. The prospective two-volume perfusion measurements were quantitatively compared to the previously validated retrospective perfusion measurements with t-test, linear regression, and Bland–Altman analysis. As a result, the pulmonary artery time-to-peak ( $${T}_{PA}$$ T PA ) was related to one-half of the contrast injection duration ( $$\frac{{T}_{Inj}}{2}$$ T Inj 2 ) by $${T}_{PA}=1.01\frac{{T}_{Inj}}{2}+1.01$$ T PA = 1.01 T Inj 2 + 1.01 (r = 0.95). The prospective two-volume perfusion measurements (PPRO) were related to the retrospective measurements (PRETRO) by PPRO = 0.87PRETRO + 0.56 (r = 0.88). The CT dose index and size-specific dose estimate of the two-volume CT technique were estimated to be 28.4 and 47.0 mGy, respectively. The optimal timing protocol can enable an accurate, low-radiation-dose two-volume dynamic CT perfusion technique.