Brazilian Journal of Pharmaceutical Sciences (Dec 2009)
Quantitation of artemether in pharmaceutical raw material and injections by high performance liquid chromatography
Abstract
The quantitation of artemether in both pharmaceutical raw material and injections was carried out by high performance liquid chromatography (HPLC) with ultraviolet detection. A Zorbax C18 column (150 x 4.6 mm; 5 μm), at 30 °C, and a mobile phase composed of acetonitrile and water (70:30), at a flow rate of 1ml/min, were used. The detection wavelength was 216 nm and the injection volume was 20 μL. The method proved to be linear (r²=0.9999), precise (RSD A quantificação de artemeter em matéria-prima farmacêutica e solução injetável foi realizada por cromatografia líquida de alta eficiência (CLAE) com detecção na região do ultravioleta. Empregou-se coluna Zorbax C18 (150 x 4.6 mm; 5 μm), mantida a 30 °C, e fase móvel composta por acetonitrila e água (70:30), com fluxo de 1 ml/min. A detecção foi realizada a 216 nm, e o volume de injeção foi 20 μl. O método se mostrou linear (r²=0,9999), preciso (DPR < 2,0% para precisão intra-dia e inter-dias) e seletivo em relação a possíveis impurezas e excipientes das amostras. Os limites de detecção e quantificação obtidos foram 8 μg/mL e 25 μg/mL, respectivamente. O teor médio de artemeter obtido na análise da matéria-prima farmacêutica foi 99,26% e na solução injetável, 102,08%. O método otimizado e validado pode ser utilizado com sucesso para análises rotineiras em controle de qualidade.
Keywords