Brazilian Journal of Pharmaceutical Sciences (Dec 2009)

Quantitation of artemether in pharmaceutical raw material and injections by high performance liquid chromatography

  • Isabela da Costa César,
  • Gerson Antônio Pianetti

DOI
https://doi.org/10.1590/S1984-82502009000400018
Journal volume & issue
Vol. 45, no. 4
pp. 737 – 742

Abstract

Read online

The quantitation of artemether in both pharmaceutical raw material and injections was carried out by high performance liquid chromatography (HPLC) with ultraviolet detection. A Zorbax C18 column (150 x 4.6 mm; 5 μm), at 30 °C, and a mobile phase composed of acetonitrile and water (70:30), at a flow rate of 1ml/min, were used. The detection wavelength was 216 nm and the injection volume was 20 μL. The method proved to be linear (r²=0.9999), precise (RSD A quantificação de artemeter em matéria-prima farmacêutica e solução injetável foi realizada por cromatografia líquida de alta eficiência (CLAE) com detecção na região do ultravioleta. Empregou-se coluna Zorbax C18 (150 x 4.6 mm; 5 μm), mantida a 30 °C, e fase móvel composta por acetonitrila e água (70:30), com fluxo de 1 ml/min. A detecção foi realizada a 216 nm, e o volume de injeção foi 20 μl. O método se mostrou linear (r²=0,9999), preciso (DPR < 2,0% para precisão intra-dia e inter-dias) e seletivo em relação a possíveis impurezas e excipientes das amostras. Os limites de detecção e quantificação obtidos foram 8 μg/mL e 25 μg/mL, respectivamente. O teor médio de artemeter obtido na análise da matéria-prima farmacêutica foi 99,26% e na solução injetável, 102,08%. O método otimizado e validado pode ser utilizado com sucesso para análises rotineiras em controle de qualidade.

Keywords