Telescoping catheter technique in percutaneous coronary intervention. Case resolution

  • José Ramón Rumoroso Cuevas,
  • Mario Sádaba Sagredo,
  • Asier Subinas Elorriaga

DOI
https://doi.org/10.24875/RECICE.M21000250
Journal volume & issue
Vol. 4, no. 1
pp. 75 – 77

Abstract

Read online

CASE RESOLUTION A scheduled procedure was decided. An 8-Fr 45 cm long Flexor Shuttle-SL introducer guiding sheath (Cook, United States) was inserted via femoral access again. An 8-Fr AL2 guide catheter was inserted into the ostium of the aortocoronary graft with a valve with a body shorter than usual, the FLO40XR hemostasis valve (Merit, United States). Afterwards, a 120 cm straight tip 5-Fr Heartrail ST01 guide catheter was inserted (Terumo, Japan) that was connected to a different FLO40XR valve. The SION blue Extra Support guidewire (Asahi Intecc, Japan) was advanced towards the left circumflex artery, and mounted over it, a 5-Fr GuideLiner guide extension catheter (Teleflex, United States) was advanced. The goal of this telescoping catheter system was to increase support to be able to move the stent forward (figure 1 and figure 2). Figure 1. Arrangement of the catheters with their special valves to allow a longer usable length. The valves with the shortest bodies allow longer effective catheter lengths to reach the target lesion. Figure 2. Quadruple telescoping catheter system mounted over the SION blue ES guidewire. The compatibility between the 5-Fr guide extension catheter and the 5-Fr ST01 guide catheter is seen...