European Journal of Hybrid Imaging (Sep 2023)

Lung CT stabilization with high-frequency non-invasive ventilation (HF-NIV) and breath-hold (BH) in lung nodule assessment by PET/CT

  • Mario Jreige,
  • Emeline Darçot,
  • Alban Lovis,
  • Julien Simons,
  • Marie Nicod-Lalonde,
  • Niklaus Schaefer,
  • Flore Buela,
  • Olivier Long,
  • Catherine Beigelman-Aubry,
  • John O. Prior

DOI
https://doi.org/10.1186/s41824-023-00175-4
Journal volume & issue
Vol. 7, no. 1
pp. 1 – 10

Abstract

Read online

Abstract Purpose To evaluate the effect of lung stabilization using high-frequency non-invasive ventilation (HF-NIV) and breath-hold (BH) techniques on lung nodule detection and texture assessment in PET/CT compared to a free-breathing (FB) standard lung CT acquisition in PET/CT. Materials and methods Six patients aged 65 ± 7 years, addressed for initial assessment of at least one suspicious lung nodule with 18F-FDG PET/CT, underwent three consecutive lung PET/CT acquisitions with FB, HF-NIV and BH. Lung nodules were assessed on all three CT acquisitions of the PET/CT and characterized for any size, volume and solid/sub-solid nature. Results BH detected a significantly higher number of nodules (n = 422) compared to HF-NIV (n = 368) and FB (n = 191) (p < 0.001). The mean nodule size (mm) was 2.4 ± 2.1, 2.6 ± 1.9 and 3.2 ± 2.4 in BH, HF-NIV and FB, respectively, for long axis and 1.5 ± 1.3, 1.6 ± 1.2 and 2.1 ± 1.7 in BH, HF-NIV and FB, respectively, for short axis. Long- and short-axis diameters were significantly different between BH and FB (p < 0.001) and between HF-NIV and FB (p < 0.001 and p = 0.008), but not between BH and HF-NIV. A trend for higher volume was shown in FB compared to BH (p = 0.055) and HF-NIV (p = 0.068) without significant difference between BH and HF-NIV (p = 1). We found a significant difference in detectability of sub-solid nodules between the three acquisitions, with BH showing a higher number of sub-solid nodules (n = 128) compared to HF-NIV (n = 72) and FB (n = 44) (p = 0.002). Conclusion We observed a higher detection rate of pulmonary nodules on CT under BH or HF-NIV conditions applied to PET/CT than with FB. BH and HF-NIV demonstrated comparable texture assessment and performed better than FB in assessing size and volume. BH showed a better performance for detecting sub-solid nodules compared to HF-NIV and FB. The addition of BH or HF-NIV to PET/CT can help improve the detection and texture characterization of lung nodules by CT, therefore improving the accuracy of oncological lung disease assessment. The ease of use of BH and its added value should prompt its use in routine practice.

Keywords