Diagnostics (Apr 2022)

Programming Optimization in Implantable Cardiac Monitors to Reduce False-Positive Arrhythmia Alerts: A Call for Research

  • Fabrizio Guarracini,
  • Martina Testolina,
  • Daniele Giacopelli,
  • Marta Martin,
  • Francesco Triglione,
  • Alessio Coser,
  • Silvia Quintarelli,
  • Roberto Bonmassari,
  • Massimiliano Marini

DOI
https://doi.org/10.3390/diagnostics12040994
Journal volume & issue
Vol. 12, no. 4
p. 994

Abstract

Read online

No studies have investigated whether optimizing implantable cardiac monitors (ICM) programming can reduce false-positive (FP) alerts. We identified patients implanted with an ICM (BIOMONITOR III) who had more than 10 FP alerts in a 1-month retrospective period. Uniform adjustments of settings were performed based on the mechanism of FP triggers and assessed at 1 month. Eight patients (mean age 57.5 ± 23.2 years; 37% female) were enrolled. In 4 patients, FPs were caused by undersensing of low-amplitude premature ventricular contractions (PVCs). No further false bradycardia was observed with a more aggressive decay of the dynamic sensing threshold. Furthermore, false atrial fibrillation (AF) alerts decreased in 2 of 3 patients. Two patients had undersensing of R waves after high-amplitude PVCs; false bradycardia episodes disappeared or were significantly reduced by limiting the initial value of the sensing threshold. Finally, the presence of atrial ectopic activity or irregular sinus rhythm generated false alerts of AF in 2 patients that were reduced by increasing the R-R variability limit and the confirmation time. In conclusion, adjustments to nominal settings can reduce the number of FP episodes in ICM patients. More research is needed to provide practical recommendations and assess the value of extended ICM programmability.

Keywords