CoDAS (Apr 2022)

Can the use of the CROS system provide head shadow effect contribution to unilateral Cochlear Implant Users?

  • Ana Cristina Hiromi Hoshino,
  • Maria Valéria Schmidt Goffi-Gomez,
  • Paola Angelica Samuel Sierra,
  • Smita Agrawal,
  • Carina Rodriguez,
  • Ana Claudia Martinho de Carvalho,
  • Robinson Koji Tsuji

DOI
https://doi.org/10.1590/2317-1782/20212021071
Journal volume & issue
Vol. 34, no. 5

Abstract

Read online

ABSTRACT Purpose The aim of this study was to evaluate the contribution of the CROS system on the head shadow effect in unilateral implant users. Methods Prospective cross-sectional study, approved by the ethics committee under protocol 2.128.869. Eleven adults with post-lingual deafness users of unilateral Advanced Bionics CI were selected. Speech recognition was evaluated with recorded words presented at 65dBA at 0o azimuth and at 90o on the side contralateral to the CI, with noise at 55dBA, using CI alone and CI + CROS system. The results were analyzed using paired t-test with a 0.05 alpha. Results The mean speech recognition scores were significantly better with CI + CROS in relation to the condition of CI alone (p <0.05, p <0.005 and p <0.005 respectively). In the presentation at 0o azimuth, no significant differences were found. Conclusion Users of unilateral CI without useful residual hearing for the use of hearing aids or unable to undergo bilateral surgery can benefit from the CROS device for speech recognition, especially when the speech is presented on the side contralateral to the CI.

Keywords